Model PB-5 | How do I run MCA linearity measurements?

Question 

Model PB-5 | How do I run MCA linearity measurements?

Answer 

Best results are accomplished by multiple sweeps at the shortest ramp time (90 seconds). Set the number of sweeps to 999, giving a run time of just over 24 hours. For unusually demanding needs, the measurement can be repeated for several days. Since any temperature corrections are made between sweeps (not during the ramp), ramp times of 15 minutes may not give the required accuracy. Short ramps allow good random statistics for all channels as long as many sweeps are made.

FAQ Category 
Test & Measurement