  • Trigger Internal 0.01 Hz to 100 MHz; external 0 to 100 MHz
  • Pulse Width 3 ns to 640 s
  • Impulse 400 ps typical (fixed)
  • Delay 0 ns to 640 s (plus fixed delay)
Applications
  • Photo Multiplier
  • Photo Detector
  • Fiber Optics Testing
Specifications 

LIGHT OUT

Spectral Peak 650nm / 850nm / 904nm  (+/- 15nm typical contact factory for selected wavelengths)
Spectral Width less than 5nm (nominal)
Power Levels 0 to 2 mW baseline, 100 uW to 2 mW peak; independently adjustable. Impulse output fixed at 100 uW baseline, 1.5 mW peak (typical).
Custom Peak Power 065 - 10 mW | 085 - 20 mW | 095 - 20mW
Resolution 10 uW, 100 uW to 2 mW pulses; 0 to 2 mW in External Modulation and CW.
Transition Times (20% to 80%) 500 ps rise time, 1.5 ns fall time (pulses with zero baseline) 1 ns rise time, 1 ns fall time (pulses with non zero baseline).
Dynamic Range 13 dB, pulses with zero baseline; 20 dB, pulses with nonzero baselineExternal Modulation and CW.
Accuracy (Nonzero Baseline) Absolute: +/- 1 dB, 200 uW to 2 mW. Relative: +/- 0.5 dB, 200 uW to 2 mW pulses;+ 30 uW, 200 uW to 2 mW for External Modulation and CW.
Temperature Coefficient 0.05 dB/°C.
Fiber 6/125 Single Mode Fiber standard (Contact Factory for custom); 0.21 NA.
Connector ST Standared (Contact Factory for custom)

TIMING

Trigger Internal 0.01 Hz to 100 MHz; external 0 to 100 MHz.
Pulse Width 3 ns to 640 s.
Impulse 400 ps typical (fixed).
Delay 0 ns to 640 s (plus fixed delay).
External Drive 0 to 200 MHz, pulses with zero baseline; 0 to 300 MHz, pulses with nonzerobaseline; 300 mV min. amplitude; 50 ohm input.
External Modulation 100 Hz to 700 MHz (-3 dB); 100mV rms/mW sensitivity (typical); 50 ohminput.
Module Delay 10 ns typical (fixed) mainframe's Pulse Out to Light Out. Jitter 100ps rms.

STORE

A ROM determines the instrument's allowable operating conditions and units of display applicable for the installed module.

A nonvolatile memory (RAM) in each module stores up to ten complete settings of operating conditions (trigger level, slope, etc.) and output pulse parameters.

RECALL

The RAM-stored settings can be manually or bus transferred to the mainframe to produce the predetermined outputs.

WEIGHT

2 Ibs net (0.9 kg); 5 Ibs shipping (2.3 kg). (When installed in a Model 6040 Mainframe, est. 24 lbs. shipping)

AMBIENT

Same as Model 6040 mainframe.

Downloadable Resources 

Downloadable resources such as datasheets, firmware, software, drivers and products manuals. Alternatively, you can browse resources directly by visiting our downloads page.

Product Datasheets
• Product Firmware
• Product Software and Drivers
• Product Manuals

Price List

Price lists are available to our registered users. To view pricing for this and other products, please log in or create a free account.